Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning After An Overflow; Putting Into Operation After Cleaning; Technical Data General Data; General Data - Mettler Toledo XPR26PC Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für XPR26PC:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
Cleaning around the balance
Remove any dirt or dust around the balance and avoid further contaminations.
Cleaning the terminal
Clean the terminal with a damp cloth or a tissue and a mild cleaning agent.
Cleaning the removable parts
Clean the removed part with a damp cloth or a tissue and a mild cleaning agent.
Cleaning the weighing unit
1 Disconnect the balance from the AC/DC adapter.
2 Use a lint-free cloth moistened with a mild cleaning agent to clean the surface of the balance.
3 Remove powder or dust with a disposable tissue first.
4 Remove sticky substances with a damp lint-free cloth and a mild solvent, e.g., isopropanol or ethanol
70%.

5.3.3 Cleaning after an overflow

In the event of an overflow, for example when overfilling the pipetting container, excess liquid must be removed
immediately.
1 Use the pump to empty the evaporation trap container, see [Emptying the evaporation trap container 
 Page 19].
2 Use the pump to empty the pipetting container, see [Emptying the pipetting container   Page 18].
3 Disassemble the remaining items and clean them with a lint-free cloth or tissue, see [Disassembling for
cleaning   Page 20].
4 Reassemble the balance.
See also
Assembling the pump   Page 18

5.3.4 Putting into operation after cleaning

1 Reassemble the balance.
2 Check if the terminal is connected to the balance.
3 Reconnect the balance to the AC/DC adapter.
4 Check the level status, level the balance if necessary.
5 Respect the warm-up time specified in the "Technical Data".
6 Perform an internal adjustment.
7 Perform a routine test according to the internal regulations of your company. METTLER TOLEDO
recommends to perform a sensitivity test after cleaning the balance.
8 Press
The balance is ready to be used.
See also
Performing an internal adjustment   Page 14
6 Technical Data

6.1 General data

Power supply
AC/DC adapter (model no.
FSP060‑DHAN3):
Pipette Calibration Balances
to zero the balance.
Input: 100 – 240 V AC ± 10%, 50 – 60 Hz, 1.8 A
Output: 12 V DC, 5 A, LPS, SELV
u
www.mt.com/lab-cleaning-guide
21

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis