RID16
Abb. 5: Mounting the cable gland or fieldbus connector
A
Mounting plate
B
Cable gland
C
Fieldbus connector
1.
First of all using a suitable tool, such as a screwdriver, open one of the indentations pro-
vided on the bottom of the device. Open the indentation at room temperature, as otherwise
the housing could be damaged at very low temperatures.
2.
Install the mounting plate for the cable gland and fieldbus connector. The mounting plate
is supplied with the field indicator, see the scope of delivery.
3.
Insert the cable gland or the fieldbus connector into the mounting plate. A cable gland is
supplied with the field indicator, see the scope of delivery. The fieldbus connector is avail-
able as an accessory.
Mounting the cable gland or fieldbus connector, aluminum housing
In the case of the aluminum housing, the cable gland or the fieldbus connector can be screwed
into the housing directly. A mounting plate is not required.
Endress+Hauser
Wiring
A0021557
33