5.1.3 SETUP LEVEL 3: Setting of the initial parameters for "PEG mode"

A triple-click of the SETUP button will activate a program option which allows you to individually preset
the stimulation parameters for the PEG-handle after the device is switched on. The parameters and the
manufacturer's settings are listed below.
Depressing the SETUP button for a longer period of time (min. 2 seconds) will permanently store the
values currently indicated on the display. A monitoring sound is audible during the store-procedure,
then you are redirected to the PAUSE-mode.
The settings for the frequency and the impulse bandwidth must be adjusted in normal operation. i.e.
before pressing the SETUP button for the first time. The intensity of the current can still be varied in the
SETUP-function with the aid of the turn-knob.
Initial parameters for stimulation using the PEG-handle
PEG parameter
Stimulation current intensity
Stimulation frequency
Stimulation impulse bandwidth

6. Error messages

After switching the MultiStim SENSOR on, the device will automatically carry out a self-test. If the
device recognises a faulty function during the self-test or during normal operation, the corresponding
error code will be indicated on the LCD.
After that, the device is not operational anymore and may not be used further.
The device must by all means be returned to the manufacturer for repair. Please ask your dealer if you
need assistance, he will be glad to help you.
Please indicate the error code in the repair order. The following error codes are possible:
E1 Fault in Program memory
E3 Faulty pulse voltage
E4 Faulty internal current setting
E5 Faulty pulse current (perhaps damaged cable)
E6 Battery is empty
If you see any error codes that are not listed here, please contact the manufacturer or your dealer.

7. The battery

The proper charge of the battery must be examined regularly. Proceed as described in section Indica-
tion of battery condition. A battery capacity of less than 20 % (1 segment visible) will cause the bat-
tery condition indicator to start flashing, and the battery should be replaced. If the battery capacity
drops to a value which will impair the functionality, the device will switch itself off automatically. If the
MultiStim-device is not used for a longer period of time, the battery must be removed to avoid leakage.

7.1 Battery replacement

The battery compartment is located on the bottom side of the MultiStim SENSOR. Open the compart-
ment by unscrewing the screw situated on the back side of the device with a suitable tool. Replace the
battery. Be sure to check for correct polarity. Exclusively use 9 V alkali-manganese batteries (e.g., VARTA
4022, DURACELL MN 1604). With these batteries you will obtain a prolonged duration of functionality
and extremely reliable operation.
Attention:
In case of battery leakage, the device should not be operated anymore for safety reasons. If acid leaks
into the inside of the device, essential functional assemblies may be damaged or impaired. The device
must be returned to the manufacturer for inspection.

8. Cleaning and disinfection of the device

Cleaning can be carried out by wiping with disinfectant. Take care that no water or moisture enters
into the device. Alcohol, or commercially available alcohol based disinfectants containing no methyl
alcohol may be used for disinfection.
54
XS190056I.indb 54
Adjustment range
0 – 60,00 mA
1 Hz, 2 Hz
0,05 ms/ 0,1 ms/ 0,2 ms/
0,3 ms/ 0,5 ms/ 1,00 ms
Manufacturer's settings
10 mA
1 Hz
0,5 ms
08.12.17 09:33
At
Th
tri
ne
9.
Ch
at
au
9.
Th
tio
9.
Th
Th
1.
2.
3.
4.
Th
10
Th
th
11
Fo
en
pa
pe
th
Pu
Fr
Br

Werbung

Inhaltsverzeichnis
loading

Verwandte Produkte für pajunk MultiStim SENSOR

Inhaltsverzeichnis