Symbol
Meaning
Indication of the battery charge level
Each of the five segments represents 20 % of the battery capacity. This means: if all
five segments are visible, then 100 % of the utilisable battery capacity are available.
Operating modes: PEG, cannula.
Indicates, that setup mode is activated.
Indication of impulse bandwidth and the frequency. The numerals and units shown
within the symbol render the values selected with the control keys (see description).

3.5 Menu structure setup

SETUP level 1
Adjustment/ storage of the volume
Options to choose from:
SETUP level 2
Setting the starting parameters for the cannula mode
Options to choose from:
SETUP level 3
Setting the starting parameters for the PEG mode
Options to choose from:

4. Operation

4.1 Pre-operational check

Please observe: Equipment with divergent behaviour may not be put into operation. In this case,
please contact the customer service. Electro-medical devices may only be repaired by the manufacturer
or by an institution expressly authorised by the manufacturer.
Please go through the following checking procedure before the initial startup:
1. Press the ON button to start the device. At this moment, the device will automatically start a self-test
sequence. After successful completion of the self-test sequence, the device will switch to the PAUSE
mode. The LCD display will inform you about the current settings. Please replace the battery imme-
diately if there is no indication visible after the device has been switched on. (Observe section Bat-
tery.) If the self-test sequence should perhaps have identified a faulty function, then the correspond-
ing error code will be indicated on the LCD. After this occurrence, the device is not operational any
more. (Observe section Error Messages.)
2. Check the electrode cable by visual inspection. Damaged cables may not be used. Attach the elec-
trode cable at the top end of the MultiStim device as follows: Place the electrode cable plug in the
device sleeve so that the groove on the plug sleeve is aligned with the groove and the red marking
on the device sleeve.
The plug can be plugged in only in this position; it may not be
inserted in any case by using force. Now insert the plug all the
way. When the plug has been inserted, it is locked in position and
can not be inadvertently disconnected by pulling at the electrode
cable. To disconnect the electrode plug, take hold of it at the
structured surface near the red dot using your thumb and index finger. The locking mechanism of
XS190056I.indb 49
A4, A3, A2, A1
(decreasing volume, beep sounds)
0
(global mute)
1, 2, 3, 4
(increasing volume, pulse sounds)
set value (control device) storage
set value (control device) storage
Abb.1
Abb. 2