Guidance and Manufacturer's Declaration - Electromagnetic Emissions
(according to EN 60601-1-2:2007; 5.2.2.1 Table 2)
MultiStim SENSOR is intended for use in the electromagnetic environment specified below. The
customer or user of the MultiStim SENSOR should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
Electrical fast tran-
sient/ burst
according to
IEC 61000-4-4
Surges according to
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines accord-
ing to
IEC 61000-4-11
Power frequency
(50/ 60 Hz) mag-
netic field accord-
ing to
IEC 61000-4-8
U
is the a.c mains voltage prior to application of the test level.
r
38
XS190056I.indb 38
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
< 5 % U
r
(>95 % dip in U
)
r
for 0,5 cycle
40 % U
r
(60 % dip in U
)
r
for 5 cycles
70 % U
r
(30 % dip in U
)
r
for 25 cycles
< 5 % U
r
(>95 % dip in U
)
r
for 5 s
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
Not applicable
Not applicable
Not applicable
3 A/m
Electromagnetic envi-
ronment – guidance
Floors should be wood,
concrete or ceramic
tile. If floors are covered
with synthetic material,
the relative humidity
should be at least 30 %.
Not applicable
Not applicable
Not applicable
Power frequency mag-
netic fields should be
at levels characteristic
of a typical location in
a typical cornmercial or
hospital environment.