Guidance and Manufacturer's Declaration - Electromagnetic Emissions
(according to EN 60601-1-2:2007; 5.2.2.2 Table 3)
MultiStim SENSOR is intended for use in the electromagnetic environment specified below. The
customer or user of the MultiStim SENSOR should assure that it is used in such an environment.
Immunity test
Conducted RF
according to
d
IEC 61000-4-6
.
Radiated RF
according to
IEC 61000-4-3
Note 1
XS190056I.indb 39
IEC 60601 test level
Compliance level
Not applicable
3 V
rms
150 kHz to 80 MHz
outside the ISM-bands
a
10 V
Not applicable
rms
150 kHz to 80 MHz in
the ISM-bands
a
10 V/m
10 V/m
80 MHz to 2.5 GHz
At 80 MHz and 800 MHz, the higher frequency range applies.
Electromagnetic envi-
ronment - guidance
Portable and mobile
RF communications
equipment should not
be used no closer to
any part of the Multi-
Stim SENSOR, includ-
ing cables, than the
recommended separa-
tion distance calculated
from the equation
applicable to frequency
of the transmitter.
Recommended separa-
tion distance:
d = 3,5/U1√P for
150 kHz to 80 MHz
d = 12/U1√P for
80 MHz to 800 MHz
d = 23/U1√P for
800 MHz to 2,5 GHz
where P is the maxi-
mum output power
rating of the trans-
mitter in watts (W)
according to the trans-
mitter manufacturer
and d is the recom-
mended separation dis-
tance in meters (m)
.
b
Field strengths from
fixes RF transmitters,
as determined by an
electromagnetic site
survey,
should be less
c
than the compliance
level in each frequency
range.
d
Interference may occur
in the vicinity of equip-
ment marked with the
following symbol:
39