5. SETUP-functions

5.1.1 SETUP LEVEL 1: Volume control

s
to
Pressing the SETUP button once will activate a program option which allows you to control the volume
).
of the monitoring sound as well as of the warning sounds.
Your display will show a number appearing on the left-hand side next to the VOL. indication (see
table). This number corresponds to a certain predefined volume.
This volume may be increased or reduced with the aid of the turning-knob. A value of 4 corresponds
to maximum volume. Turning the value to 0 will shut the monitoring sound off. Depressing the SETUP
button for a longer period of time (min. 2 seconds) will permanently store the volume-value currently
indicated on the display. A monitoring sound is audible during the store-procedure.
The device will automatically exit the SETUP-mode and switch to the PAUSE-mode. A short depression
of the SETUP button will guide you to the next item of the SETUP-menu without having stored the
volume-settings.
Volume of the monitoring- and warning sounds
Display
4
3
2
1
0
A1
A2
A3
A4

5.1.2 SETUP LEVEL 2: Setting of the initial parameters for "Cannula mode"

Double-clicking the SETUP button will activate a program option which allows you to individually
nt
preset the stimulation parameters for the cannula after the device is switched on. The parameters and
e
the manufacturer's settings are listed below.
Depressing the SETUP button for a longer period of time (min. 2 seconds) will permanently store the
values currently indicated on the display. A monitoring sound is audible during the store-procedure,
then you are redirected to the PAUSE-mode. A short depression of the SETUP button will guide you to
the next item of the SETUP-menu without having changed the existing settings (see 3.8.3).
e
Please observe:
The settings for the frequency and the impulse bandwidth must be adjusted in normal operation. i.e.
.
before pressing the SETUP button for the first time. The intensity of the current can still be varied in the
SETUP-function with the aid of the turn-knob.
ATTENTION:
Having selected a too high initial intensity of the stimulation current may lead to painful stimulation of
the patient during subsequent introduction of the cannula.
Initial parameters for stimulation using the cannula
Cannula-parameter
Stimulation current intensity
-
Stimulation frequency
Stimulation impulse bandwidth
XS190056I.indb 53
Monitoring sounds
Warning sounds
very loud
very loud
loud
loud
medium
medium
low
low
off
off
off
low
off
medium
off
loud
off
very loud

Adjustment range

0 – 6,00 mA
1 Hz, 2 Hz
0,05 ms/ 0,1 ms/ 0,2 ms/
0,3 ms/ 0,5 ms/ 1,00 ms
Monitoring- and
warning sounds
Warning sounds only
Manufacturer's settings
1,50 mA
2 Hz
0,1 ms
53
08.12.17 09:33