8 Error codes (in the data records)

The error codes are not displayed as text in the data recording and when querying via the
interface. This has the background that with different languages otherwise no reference to
the actual error can be made.
Error code
Text
0
OK
100000000
Measuring range exceeded
100000001
Measuring range not reached
100000010
Calculation not possible
100000011
System error
100000012
Battery empty
100000013
No sensor
100000014
Recording error
100000015
EEPROM checksum wrong
100000016
System restart
100000017
Data display error
100000018
Invalid data
100000020
Recording finished
100000021
Recording started
100000022
Channel deactivated
100000023
Temperature channel deactivated
100000024
No temperature sensor
100000025
No data available
-23
Sensor module does not respond
-10
Non-existent
-255
Unexpected error
-100
Error calibration
-75
Not found
-101
Not calibrated
-253
Value not stable
-251
Not in temperature range
*If the error remains, send device to service.
58
Notes
No error
Check calibration and sensor
Check calibration and sensor
Check settings
Restart the device*
Charge device
Plug in sensor
Restart the device*
Restart the device*
Device starts, wait briefly
Restart the device*
Restart the device*
Logger was stopped
Logger was started
Channel is deactivated in the device
menu
Check setting for temperature com-
pensation, activate O
2
Connect temperature sensor, check
setting for temperature compensa-
tion, activate O
/Con input
2
Device has not measured anything
yet
Restart the device*
Restart the device*
Restart the device*
Perform calibration again
Restart the device*
Perform calibration
Ensure a stable environment
Check temperature