Remove the nut from the sensor.
4.
While the sensor is still connected to the power supply, fix it
with a fastener in the beaker.
5.
When the sensor displays a stable value, take a sample with a
pipette from the beaker and perform a DPD-1 measurement.
6.
After the calibration, put back the sensor into the analytical mea-
surement chamber.
9.4.
Troubleshooting
9.4.1.
leD signalling on the sensor with
order code 565164
One orange LED and one green LED show the status of the sensor:
LED
LED
Meaning
status
Green
Correct operation
of the sensor
ON
OFF or
Supply voltage is
flashing
too low.
22
Recommended actions
—
→
Increase the value of
the supply voltage.
english
Type 8232
Maintenanceandtroubleshooting
LED
LED
Meaning
status
Orange
Correct operation
OFF
of the sensor
The sensor is not
correctly wired.
ON
The concentration
of the chlorine in
the water is too
Flashing
high.
Recommended actions
—
→
Wire the sensor
according to chap 7.
→
Reduce the chlorine
concentration, see
chap 6.3.