Technicaldata
6.

TechnicAl DATA

6.1.

operating conditions

Ambient temperature
For sensors with order code 565164: +5 to +40°C
For sensors with order codes 566051 and 566052: +5 to +45°C
Air humidity
< 90%, without condensation
Height above sea level
Max. 2000 m
6.2.

conformity to standards and directives

The device conforms to the EC directives through the following standard: EMC: EN 61326-1.
6.3.

general technical data

Type of medium
Medium temperature
Maximum medium pressure
Medium pH range
Type 8232
566051
• swimming pool water, drinking
water, service water, process
water
• free of any surfactants
• with constant pH value.
+5 to +45 °C
1 bar, no pressure drops are
allowed
6-8 pH
566052
• swimming pool water, drinking
water, sea water
• surfactants are partially
tolerated.
+5 to +45 °C
0,5 bar, no pressure drops are
allowed
4-9 pH
english
565164
water with drinking water quality
+5 to +40 °C
0,5 bar, no pressure drops are
allowed
6,5-9 pH