566051
14
Hold the shaft upright and slowly insert it into the filled membrane cap.
Hold the membrane cap and slowly screw the shaft clockwise onto the
membrane cap. If a resistance occurs because of the seal, continue
screwing until the working electrode and the bottom of the membrane
cap are in contact: the membrane then has a convex curvature.
Make sure the shaft is tightly fastened to the membrane cap:
- do not put your fingers on the vent because excess electrolyte
must flow out through the vent;
- the membrane has a convex curvature: do not touch nor adjoin the
membrane. Do not thump the membrane else it would be damaged
and unusable.
Remove excess electrolyte with a cloth dampened with water.
Cover the vent with the hose ring.