WARNINGS
AND
1.4. Main Switch with
Emergency Off Function
The machine has a main switched combined with an
emergency off function. If the switch is turned to
OFF, the machine is immediately cut off from the
power supply network, the exposure lamp goes off,
and the upper microscope pivots upwards. The
switch is signal red and is located on the left front
plate.
Attention!I
In case of danger, turn the main switch
to OFF!
Warning!
The upper microscope pivots upwards.
To begin operating the machine again, the main
switch must be turned to ON and the machine must
be restarted.

1.5. Dangers

1.5.1. Physical Dangers

Moving machine parts such as the alignment station,
microscopes and their manipulators, the mirror
housing and others can cause injury to the operator.
You should take particular care of the movements of
the microscope if your machine is equipped with the
direct microscope view finder.
Warning!
Moving machine parts can injure the fin-
gers or face.
Do not remove any materials from the
machine until all machine movements
have stopped!
Warning!
With a machine equipped with a vibra-
tion-damped table, the base plate can
sink without warning in the event of a
drop in pressure in the compressed air
supply.
Never place the hands between the base
plate and the vibration-damped table of
the machine!
2
- Original Documentation -
SAFETY PRECAUTIONS
SAFETY-INST MJB4 Rev09 17-03 WARNINGS and SAFETY PRECAUTIONS
Warning!
The movement of a direct viewfinder
can cause injury to the eyes or face.
Stop looking into the microscope view-
finder before you start the microscope
movement!
Danger!
A particular danger results from the
fragments thrown out in the event of a
lamp explosion!
Situations which can lead to a lamp ex-
plosion are described in a separate sec-
tion.
You should also be aware of the risk of
poisoning with Hg following a lamp ex-
plosion; this is also described in a sep-
arate section.

1.5.2. Thermal Dangers

The upper side of the lamp house gets very hot dur-
ing operation.
Operating temperatures of several hundred °C occur
on the lamp itself.
The lamp cooling must be correctly adjusted accord-
ing to the lamp type!
Warning!
There is a risk of lamp explosion if the
lamp house is opened prematurely
while the lamp is hot.
The exposure lamp has to cool down for
at least 20 minutes before the lamp
house door is opened!
The controlled cooling required for
more powerful lamp types must also be
left in operation during the cooling
phase!

1.5.3. Electrical Dangers

Live parts may be exposed when the machine cover
is removed.
High wattages and currents of up to 50 A can occur
in the lamp and lamp power supply.
The lamp starting voltage can be up to 30 kV.