Table 1: Settings for Lamp Cooling and Lamp Housing Extraction
Exposure lamp
[W]
Nitrogen
[bar]
Main pressure
Nitrogen lower
[l/min]
lamp base
Nitrogen upper
[l/min]
lamp base
Nitrogen
[l/h]
Shutter
Diameter of
[mm]
exhaust on
bottom
Notes
Lamps that can be used:
•
HBO 200 W/DC - PN: G191806
(additional lamp adapter required, item no.:
G189336)
•
HBO 350 W-S - PN: G146 049
•
USH 350 DS - PN: G182 538
UXM 501 MD - PN: G148 979
(additional lamp adapter required, item no.:
G202485)
SAFETY-INST MJB4 Rev09 17-03 INSTALLATION AND TRANSPORT
200
350
Osram
Osram
1.5
1.5
2.0
3.0
-
-
260
260
closed
closed
reduced
DUV+Ozone
INSTALLATION AND TRANSPORT
350
500
Ushio
Ushio
1.5
1.5
3.0
4.7
(distributed
to both,
upper and
lower lamp
base)
-
260
260
closed
100 required
for DUV appl.
REVERSED
polarity!
- Original Documentation -
Notes
± 0,5 bar
Must be set by
service technician.
Adjustment by 1/
FM1-B10
Must be set by
service technician
needs adapter tube
outlet temp.< 60°C
11