Table 6: Recommended separation distance between portable and mobile RF communica-
tion equipment and the ME device or ME system – for ME devices or ME systems
which are not life-sustaining (refer to 5.2.2.2)
Recommended separation distances between portable and mobile RF communication equipment
and the nursing home bed "carisma".
The nursing home bed "carisma" is intended for use in an ELECTROMAGNETIC ENVIRONMENT in which radiated
RF disturbances are controlled. The client or the user of the nursing home bed "carisma" can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communica-
tion equipment (transmitters) and the nursing home bed "carisma" as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power of transmitter
[W]
0.01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is
the maximum output power of the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
3
5 ,
=
d
P
U
1
0.12
0.37
1.17
3.69
11.67
80 MHz to 800 MHz
3
5 ,
=
d
P
E
1
0.12
0.37
1.17
3.69
11.67
800 MHz to 2.5 GHz
7
=
d
P
E
1
0.23
0.74
2.33
7.38
23.33