Table 4: Guidelines and MANUFACTURER'S declaration – electromagnetic COMPATIBILITY –
for all ME DEVICES and ME SYSTEMS which are not LIFE-SUSTAINING
(refer to 5.2.2.2).
The nursing home bed "carisma" is intended for use in an ELECTROMAGNETIC ENVIRONMENT as described
below. The customer or user of the nursing home bed "carisma" must ensure that the bed is only used in a
corresponding environment.
COMPATIBILITY tests
Conducted RF
(IEC 61000-4-6)
Radiated RF
(IEC 61000-4-3)
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Field strength from fixed transmitters such as base stations for radio (cellular/cordless) telephones and land
a
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic
site survey should be considered. If the measured field strength in the location in which the nursing home bed
"carisma" is used exceeds the applicable RF compliance level above, the nursing home bed "carisma" should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as rePage 55 of 58 orienting or relocating the nursing home bed "carisma".
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
b
Page 118
IEC 60601
COMPLIANCE
TEST LEVEL
LEVEL
3 V
RMS value
150 kHz to 80 MHz
U1 = 3 V
3 V/m
E
80 MHz to 2.5 GHz
EMC GUIDELINES
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the nursing
home bed "carisma", including
cables, than the recommended
separation distance calculated
from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
=
d
=
d
RMS value
=
d
= 3 V/m
1
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance
in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
be less than the compliance level in
each frequency range.
Interference may occur
in the vicinity of equip-
ment marked with the
following symbol:
3
5 ,
P
U
1
3
5 ,
80 MHz to 800 MHz
P
E
1
7
800 MHz to 2.5 GHz
P
E
1
, should
a
b