U10M, U10S

7.4 TEDS specifications

Type
Data transfer
To switch from measurement mode to
data mode in order to transfer TEDS
data
Nominal value of supply voltage for the
TEDS module
Working range of supply voltage for the
TEDS module
Data transfer protocol
Data format

7.5 Electrical connection with TEDS

A
F
B
E
C
D
Connection with TEDS
Fig 7.1:
A1385−6.0 en/de/fr
Device plug pin assignment
B
Measurement signal (+)
D
Excitation voltage (−) (TEDS
Measurement signal (−)
C
Excitation voltage (+)
A
F
Sense lead (+)
Sense lead (−) (TEDS
E
Cable shield, connected to the housing
*)
Only with selected option T (transducer identification)
TEDS module
Apply a supply voltage between one of
the excitation voltage lines and the corre-
sponding sense lead. The sense lead
then acts as the signal lead and the ex-
citation voltage line as the ground. If this
voltage is no longer present, the module
switches back to measurement mode, in
which it no longer interacts with the ex-
citation voltage or the measurement sig-
nal.
5
V
V
3.4 ... 6.0
As specified in IEEE standard 1451.4
("one wire protocol").
As specified in IEEE standard 1451.4
Free end cable assignment
*)
)
*)
)
27
wh (white)
bk (black)
rd (red)
bu (blue)
gn (green)
gy (gray)
Shielding