Herunterladen Inhalt Inhalt Diese Seite drucken

Electrical Connection; Connection With Six-Wire Configuration - HBM U10M Montageanleitung

Kraftaufnehmer
Vorschau ausblenden Andere Handbücher für U10M:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

22
6

Electrical connection

The following can be connected for measurement signal conditioning:
• Carrier-frequency amplifier
• DC amplifier
designed for strain gage measurement systems.
The U10M/U10S force transducers are delivered with six-wire configuration
and are available with the following electrical connections:
• Bayonet connection: plug-compatible with the MIL-C-26482 series 1
connection
Connection cable KAB157-3; IP67, EMC-tested; Order number
1-KAB157-3
• Screw connection: plug-compatible with the MIL-C-26482 series 1
connection
Connection cable KAB158-3; IP64, Order number 1-KAB158-3

6.1 Connection with six-wire configuration

A
F
B
E
C
D
Fig 6.1:
Pin assignment in six-wire configuration
With this cable assignment, the output voltage at the measuring amplifier is
positive in the tensile direction when the transducer is loaded.
HBM
Device plug pin assignment
B
Measurement signal (+)
D
Excitation voltage (−)
Measurement signal (−)
C
Excitation voltage (+)
A
F
Sense lead (+)
Sense lead (−)
E
Cable shield, connected to the housing
U10M, U10S
Free end cable assignment
wh (white)
bk (black)
rd (red)
bu (blue)
gn (green)
gy (gray)
Shielding
A1385−6.0 en/de/fr

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

U10s

Inhaltsverzeichnis