InPro 6000 Aperometric O
Procedure
Cleanse with demin water; turn cap sleeve twice, do not
unmount yet. From the bottom push the membrane body out
by a slight upwards push; unscrew completely. Remove cap
sleeve and membrane body. Clean inner body. If nescessary
use new membrane body and fill with fresh electrolyte
solution. Make sure no air bubbles are trapped. Replace
O-ring each time when replacing the membrane body and
screw the cap sleeve on carefully.
7

Disposal

It is the user's responsibility to dispose of spent spare parts
and if necessary the sensor in a professional manner.
Sensors contain electronic components that require special
disposal without any hazard for humans, and with no risk of
harm to the environment. Please observe the applicable local
or national regulations concerning the disposal of "waste
electrical and electronic equipment".
8

Warranty

On manufacturing defects, 12 months after delivery.
9

Storage conditions

Inspect the sensor immediately after receiving for any
transport damage. The sensor should be stored in a dry,
clean and protected area until the time of installation. If the
sensor has been removed from the process, it should be
thoroughly cleaned and dried. It should be stored in a dry,
clean and protected area until time of installation.
10

EC Declaration of Conformity

The complete declaration is available in the declaration of
conformity certificate.