GUIDANCE AND MANUFACTURER'S DECLARATION: ELECTROMAGNETIC
SERFAS Energy Systemis intended for use in the electromagnetic environment specified below. The customer or
the user of SERFAS Energy Systemshould ensure that it is used in such an environment.
IMMUNITY TEST
Electrostatic Discharge
(ESD)
IEC61000-4-2
Electrical fast transient/
burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC61000-4-11
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
34
IMMUNITY
IEC 60601 TEST
COMPLIANCE
LEVEL
LEVEL
±6kV contact
±2,4,6kV contact
±8kV air
±2,4,8kV air
±2kV for power
±2kV for power
supply lines
supply lines
±1kV for input/
±1kV for input/output
output lines
±1kV differential
±0.5, 1kV differential
mode
±2kV common mode
±1, 2kV common
<5% Ut (>95% dip in
<5% Ut (>95% dip in
Ut) for 0.5 cycle
Ut) for 0.5 cycle
40% Ut (60% dip in
40% Ut (60% dip in
Ut) for 5 cycles
Ut) for 5 cycles
70% Ut (30% dip in
70% Ut (30% dip in
Ut) for 25 cycles
Ut) for 25 cycles
<5% Ut (>95% dip in
<5% Ut (>95% dip in
Ut) for 5 sec.
Ut) for 5 sec.
3 A/m
ELECTRO-
MAGNETIC
ENVIRONMENT:
GUIDANCE
Floors should be wood,
concrete, or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a typical
commercial or hospital
lines
environment.
Mains power quality
mode
should be that of a typical
commercial or hospital
mode
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of SERFAS Energy
Systemrequires
continued operation
during power mains
interruptions, it is
recommended that
SERFAS Energy
Systembe powered from
an uninterruptible power
supply or a battery.
Power-frequency
N/A
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.