Fore-Vacuum Pump Connection
A flange KF 40 NW (ISO 200 F inlet flange) and KF 25 NW (ISO 160 F
inlet flange) are available to connect the Turbo-V 1K-G Retrofit pump
to the fore-vacuum pump. A hose or vacuum approved pipe can be
used. If a rigid pipe is used, any vibration generated by the
mechanical pump must be eliminated through the use of bellows.
The Turbo-V 1K-G Retrofit pump is characterized by its high compression ratio
NOTE
also for oil vapors. When using a mechanical oil-sealed pump, it is advisable to
install a suitable trap between the turbopump and the fore-vacuum pump in
order to prevent oil backstreaming.