Herunterladen Diese Seite drucken

Fore-Vacuum Pump Connection - Agilent Technologies Turbo-V 1K-G Bedienungshandbuch

Vorschau ausblenden Andere Handbücher für Turbo-V 1K-G:

Werbung

Fore-Vacuum Pump Connection

A flange KF 40 NW (ISO 200 F inlet flange) and KF 25 NW (ISO 160 F
inlet flange) are available to connect the Turbo-V 1K-G pump to the
fore-vacuum pump. A hose or vacuum approved pipe can be used. If
a rigid pipe is used, any vibration generated by the mechanical pump
must be eliminated through the use of bellows.
The Turbo-V 1K-G pump is characterized by its high compression ratio also for
NOTE
oil vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.
Turbo-V 1K-G User Manual/ 87-901-025-01 (A)
Turbo-V 1K-G Navigator Controller Installation
Technical Information
293/304
17

Werbung

loading

Diese Anleitung auch für:

8698961r0028698962r001