The PROGRAMMABLE SET POINT signal has the following default
settings:
The Navigator Software (optional) allows the oper-ator to set all the
NOTE
programmable feature.
When no external input-output device is available this connector
must be closed with the supplied mating connector that short-circuits
the START and INTERLOCK inputs with the GROUND input (see the
following figure).
Figure 14
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)
Interconnections – Controller Model 9698978M004
reference quantity: frequency
threshold: 643 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second
Technical Information
MATING CONNECTOR
17
253/294