Herunterladen Inhalt Inhalt Diese Seite drucken

Bottom Mounting - Agilent Technologies Turbo-V 1K-G Bedienungshandbuch

Turbomolekularpumpen
Vorschau ausblenden Andere Handbücher für Turbo-V 1K-G:
Inhaltsverzeichnis

Werbung

17
Technical Information
Turbo-V 1K-G Retrofit Navigator Controller Installation

Bottom Mounting

See the following figure:
1
2
3
4
5
6
Figure 41
282/294
Turn the pump upside-down;
Place the three washers on the pump bottom in correspondence
of three hole at 120 °;
Place the controller on the washers, with the pump cable toward
the pump body;
Fix the controller to the pump bottom by means of the three
socket head screws M5;
Turn the pump again;
Plug the line card and connect the pump cable to the pump;
Connect the mating connector with the jumper on the interlock
signal to start the pump.
Turbo-V 1K-G Retrofit User Manual / 87-900-996-01 (B)

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis