Bottom Mounting

See the following figure.
1
2
3
4
5
6
7
Figure 56
TV 551/701 Navigator User Manual / 87-900-934-01
Turn the pump upside-down;
Place the three fixing blocks with the through holes toward the
inside, and fix them with the three M6 screws to the pump
bottom;
Place the controller on the blocks, with the pump cable toward
the pump body;
Fix the controller to the blocks by means of the three socket head
screws M5;
Turn the pump again;
Plug the line card and connect the pump cable to the pump;
Connect the mating connector with the jumper on the interlock
signal to start the pump.