PERFORMANCE SPECIFICATIONS
The table below shows ARMS (Accuracy Root Mean Square) values measured using the TC-I sensors with Masimo SET
Oximetry Technology in a clinical study.
(SpO2-SaO2) vs. (SpO2+SaO2)/2 Bland Altman fit and upper 95% and lower 95% limits of agreement.
SpO2
90-100%
80-90%
70-80%
Overall
TC-I
A
rms
2.16 (2,16) %
2.54 (2,54) %
3.97 (3,97) %
3.02 (3,02) %
70 -100%
95
9004C-eIFU-1216