Identifier meaning:
Parameter
IDD
IDO
IH1
IH2
IL1
IL2
IN1
IN2
IRT
IDA
WPP
Data transfer
ISB
GSI
GOH
GTE
GDB
TSM
ECM
ICM
Incorrect inputs, e.g. parameter limit violations or incorrect string lengths, are intercepted and
acknowledged with NAK (15H). Correct commands are acknowledged with data transfer or ACK
(06H).

4.2.8 SSI Interface

This menu is used to select the SSI parameters.
Default: GRAY 25 Bit
Settings: GRAY 24, GRAY 24+E or 25 Bit / BINARY 24 or 25 Bit

4.2.8.1 Data transfer / line length SSI

Data transfer with the "SSI" setting occurs on request from the controller; the cycle time and transfer
rate can be set within broad limits.
A clock sequence is applied to the receive input of the DME by the connected controller for this
purpose. A data bit is sent on the DME transmit line with each positive clock edge starting with the
most significant bit. A minimum interval of 30µs separates two clock sequences. The bit clock is
between 70 kHz and 1 Mhz and depends on the line length.
0000, 0050, 0100
-016383 ... +016383
0000 ... 0254
0000 ... 0254
-016383 ... +032766
-016383 ... +032766
0, 1
0, 1
020, 040, 060, 080,
100, 120, 140, 160,
180, 200
01, 02, 04, 08, 16,
32, 64, 128
0 / 1
0 , 1
Display Dimmer (%)
Display offset (mm)
Hysteresis switching output OUT1
Hysteresis switching output OUT2
Limit switching output OUT1
Limit switching output OUT2
Scaling OUT1: 0=Q, 1= Q
Scaling OUT2: 0=Q, 1= Q
Response time
Average calculation HH=00 / 1: HH=01 / 2: HH=02 /
4: HH=04 / 8: HH=08 / 16 : HH=10 / 32: HH=20 /
64: HH=40 / 128: HH=80
Write Parameter Page: stores changes to para-
meters in the EEPROM
0: normal mode, 1: Standby Note: the laser is
switched off in standby mode, no measuring values
are output
ERROR STATUS: Error status query, format:
DDDDDDDD, See Chap. 4.2.9.4 for assignment.
Operating hours (h) query format +DDDDD
Temperature (°C) query, format +DDD
ALIGN: path attenuation (dB) query, format: -DDD
Trigger Single Measurement: single measurement
trigger in mode request (see Chap. 4.2.7.3)
Execute Continuous Measurement : return to
continuous measuring value transmission after data
request
ICM 1: Request / ICM 0: Continuous
DME 3000-2 _ _