Converting Promag 50, 53 remote version to
Promag 55 remote version
8.1
Removing wall-mount housing of transmitter Promag 50, 53
CAUTION
L
The measuring device is energized!
Risk of fatal injury from electric shock.
‣
Open the measuring device only when the device is deenergized.
If, during conversion, spacing is reduced or the dielectric strength of the measuring device
cannot be guaranteed, perform a test on completion of the work (e.g. high-voltage test in
accordance with the manufacturer' s instructions).
IfPromag 50
Detach the existing S-DAT from the cable tie and remove from the transmitter.
IfPromag 53
Detach the existing S-DAT and T-DAT from the cable tie and remove from the
transmitter.