8.3
Commissioning the measuring device
• Depending on the pipe diameter and the existing mains frequency (50Hz/ 60Hz), the
measuring period and particularly the output damping must be configured in
accordance with the table in section 10. The service code is required to enter some of
the parameters.
• Complete the "Data archiving" form provided and send it to the address indicated for
archiving (data storage in the Endress + Hauser database).
• When installing a Promag 50, enter data from parameter list.
• When installing a Promag 53, use the T-DAT SAVE/LOAD function → LOAD option.
• If necessary, perform an empty pipe/full pipe adjustment as described in the Operating
Instructions.
For detailed information on the electrical connection, see the "Electrical connection"
section of the Operating Instructions for the device.
For detailed information on commissioning, see the "Commissioning" section of the
Operating Instructions for the device.
8.4
Modifying the label
Following conversion, enter the date of the conversion on the information label and attach
the information label (1) in accordance with the diagram.