5.2 Connecting to a PC

The CX22‐W Data Recorder is connected to the PC via Ethernet. There are
different variants, depending on your PC/network configuration, and the con­
nection procedure depends on the variant:
1.
CX22‐W and PC are connected directly via an Ethernet cable.
2.
The CX22‐W Data Recorder is connected with an Ethernet network
(generally automatic assignment of network configuration via DHCP,
Dynamic Host Configuration Protocol).
3.
The CX22‐W Data Recorder is connected via WLAN (CX22‐W only).
The initial configuration of a CX22‐W with WLAN must be implemented either
with monitor and keyboard (see Section 5.1) or via a connection with Ethernet
cable. You can then also activate the WLAN interface during this configuration
and enter the parameters for a WLAN connection.
Important
Wait approx. 1 minute after switching on the CX22‐W before setting up a con­
nection so that the boot process is complete and the module is ready. Other­
wise you may receive an error message that the device could not be found.
To facilitate setting up a CX22‐W connection with unknown settings, you can
use the HBM program "QuantumX Data Recorder Assistant". The settings
required are described in the following chapter. If you already know the IP
address of the CX22‐W, you can also directly set up a Remote Desktop Con­
nection (standard program in the Windows operating system). You can find
the Remote Desktop Connection program under All programs -> Accessor­
ies (Windows XP: Programs -> Accessories).

5.3 Connecting QuantumX modules

Important
If you are already using QuantumX modules and want to connect them with a
CX22‐W Data Recorder, install the latest module firmware. If you want to add
new modules at a later date, please install the latest software on the CX22‐W
using the Recovery function, see Section 11.2 on Page 39. The latest soft­
ware is always installed during restoring. You may receive a request, after
®
starting catman
A3169-1.0 en/de
AP, to update the firmware of older QuantumX modules.