6 Operating
P I
62
I6
3.
36
493
1.
2. / 5.

Anomaly Button A

The Anomaly Button A is for manual compensation of
anatomical relationships in the case of anomalies of the
anterior teeth with the P1, P2, P10, P11, P14 and P15
programs.
In case of retrusion:
move one step in + direction
In case of protrusion:
move a step or two in – direction
The setting is shown by means of an LED.
After the rotary unit returns to its original position follow-
ing the exposure, the settings automatically return to
those of normal jaws (2nd LED from the left).
i
NOTE
It is possible to set the Anomaly step, which as a rule is
set to "0" (neutral) before every exposure, to another
standard value (+1; -1; -2).
Proceed as follows:
1.
Select program P1.
For the basic Anomaly setting, the 2nd LED from the left
lights up.
2.
Press the Test Rotation key
The LED above the T key lights up.
3.
Press the Memory key.
The LED above the Memory key begins flashing.
4.
On the control panel at the side, select the required
new standard Anomaly step.
About 5 seconds after letting go of the Anomaly key, the
LED for the pre-selected Anomaly step flashes for about
1 second.
The Anomaly LED then remains lit up. The Anomaly step
is stored.
5.
Press the Test Rotation key
Sirona Dental Systems GmbH
.