Herunterladen Inhalt Inhalt Diese Seite drucken

P9 Program; P10 Program - Sirona Orthophos Plus Gebrauchsanweisung

Panorama-röntgengerät
Vorschau ausblenden Andere Handbücher für Orthophos Plus:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5 Exposure Programs
5.9

P9 Program

A
B
C
C
B
A
5.10

P10 Program

20
C
B
A
1
Film PAN
15x30 cm
6x12 in.
PAN
SID = 49.7 cm
Headpositioner
Film PAN
2
15x30 cm
6x12 in.
12,7x30,5 cm
5x12 in.
STATUS
SID = 49.7 cm
yellow
Multi-layer exposures, posterior/anterior exposures
of the temporomandibular joints
(6 exposures on 1 film)
9
20
1
Insert head positioner (see page 24)
Working area forehead support 00.0-27.0
Recommended position of forehead support 20.0
To avoid superimpositions to a large extent, head in-
clination relative to FH toward anterior.
Standard exposure (Status),
preferred mode for children
10
2
IMPORTANT!
Adjust diaphragm 2.
Yellow bite block or contact segment.
Working area forehead support 00.0− 30.0
Head inclination with the aid of FH.
Automatic selection of section shape via temple
support setting for different mandibular arches.
Radiation time dependent on adjusted width of tem-
ple supports.
Manual compensation
for anatomical relationships (protrusion, retrusion)
by anomaly button A.(see page 36)).
Sirona Dental Systems GmbH
60 82 023 D3297
D3297.201.01.11.02