5 Exposure Programs
5.5

P5 Program

5.6

P6.1 / P6.2 Program

P6.1
P6.2
18
Film PAN
15x30 cm
6x12 in.
SID = 49.7 cm
Headpositioner
P6.2
P6.1
Film PAN
15x30 cm
6x12 in.
SID = 49.7 cm
P6.1
Headpositioner
P6.2
Posterior/anterior exposures of the temporomandib-
ular joints
5
1
Insert head positioner (see page 24)
Working area forehead support 00.0− 27.0
Recommended position of forehead support 20.0
To avoid superimpositions to a large extent, head in-
clination relative to FH toward anterior.
1
PAN
Lateral exposures of the temporo-mandibular joints
with closed and open mouth
(4 exposures on 1 film)
6
1
Insert head positioner (see page 24).
Working area forehead support 00.0− 30.0
Recommended position of forehead support 06.0
Head inclination with the aid of FH.
1
P6.1 Outer Image:
Geschlossener Mund
PAN
P6.2 Inner Image:
Open mouth
Actuate P6.1
After P6.1 is completed, the unit automatically re-
turns to the initial position.
Have the patient open his mouth and actuate P6.2.