List of Error Messages
E... error messages
E1 01
One of the keys on the Multitimer was pressed
during start-up of the unit.
E1 03
Faulty communication with the unit.
E2 01
Appears after pressing exposure key.
X-ray head overheated.
Cool-down periods ignored.
E2 03
Faulty communication with the Multitimer.
E2 04
Zero-Power was re-initialized.
E2 10
Max. radiation time of program exceeded.
E2 20
Appears after pressing the exposure key,
e.g. if x-ray room door contact not closed.
Exposure lead in Multitimer cable damaged.
E2 35
Invalid data in data memory.
E2 48
Custom value for kV increase no longer set.
E3 05
Movement of film cassette obstructed.
E3 06
E3 09
Movement of height adjustment obstructed.
E3 10
Invalid stand height value displayed.
E3 11
E3 12
HV button (control panel A) was actuated during power-
up or before end of self-adjustment routine.
E3 21
Anomaly key A on control panel A was actuated during a
start-up or before selftest of the unit finished.
44
Remedy
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display. Freely pro-
grammed values (see page 42) are canceled, sorry.
Please reprogram (see page 39).
Press R key on Multitimer to reset display.
Close x-ray room door.
Press R key on Multitimer to reset display.
If error re-appears, poss. break in cable.
Press R key on Multitimer to reset display.
Have service engineer set the specific values again.
(form for fixing value)
Press R key on Multitimer to reset display.
Check that cassette can move freely.
Press R key on Multitimer to reset display.
Check that height adjustment can move freely.
Press R key on Multitimer to reset display.
WARNING
If this fault message occurs repeatedly during the
motor-driven up and down movement of the rotary
unit, especially when no patient is positioned,
switch off the unit immediately and inform your
service engineer without delay.
Switch unit off and then on again and move it in the
enabled direction until a valid display value is indi-
cated.
Press R key on Multitimer to reset display.
Press R key on Multitimer to reset display.