A reduced vacuum indicates contamination. In case of contamination, the system has to be
flushed to clean it. This flushes the vacuum pump and valves. The flushing process can also help
if the system is plugged.
Required materials:
>
Vessel with 30 ml to 60 ml distilled water, one collecting vessel
Preparations required:
The prosthesis is not worn for this process
1) Pour the distilled water into the prosthetic socket.
2) Position the vessel under the outlet hose.
3) Flex and extend the prosthetic knee joint to pump the water through the system.
4) Continue performing these movements until the full volume of water has been pumped out of
the system.
5) Clean the interior surface of the socket.
7 Cleaning
CAUTION
Use of unsuitable cleaning agents or disinfectants
Impairment of functionality and damage due to incorrect cleaning agents or disinfectants
Only clean the product with the approved cleaning agents.
Only disinfect the product with the approved disinfectants.
Observe the instructions for cleaning and care.
Inform the patient.
1) Clean the product with a damp, soft cloth.
2) Dry the product with a soft cloth.
3) Allow to air dry in order to remove residual moisture.
8 Maintenance
► The prosthetic components should be inspected after the first 30 days of use.
► Inspect the entire prosthesis for wear during normal consultations.
► Conduct annual safety inspections.

8.1 Checking for leaks

Remove the hose between the vacuum connector of the prosthetic socket and the suction connector of the
prosthetic knee joint. Use the T-piece to connect the vacuum gauge to the vacuum connector and suction con­
nector. Cover the entire vacuum connector on the inside of the socket with a small piece of paper and airtight
tape (e.g. Coroplast). Flex and extend the prosthetic knee joint until a constant vacuum of at least 300 mbar is
reached. Check whether the vacuum level is maintained.
Problem: Proximal seal
Solution: Check the proximal seal; seal or replace as
necessary.
24 | Ottobock
Remove the adhesive tape and piece of paper. Cover
the hole in the vacuum connector with a smaller piece
of paper and tape. Flex and extend the prosthetic
knee joint until a vacuum of at least 300 mbar is
reached. Check whether the vacuum level is main­
tained.