2) INFORMATION: Reducing the pretension increases knee safety while increasing the
pretension decreases knee safety.
Adjust the pretension of the EBS unit in order to achieve the desired stance phase flexion
behaviour (see Fig. 10). Ensure that the pivoting mount does not permanently touch the end
stop while walking.
Situation
Strong
stance
phase
patient experiences yielding until
the pivoting mount reaches the end
stop.
Little stance phase flexion, yielding
is not sufficient for the patient.
6 Use
CAUTION
Overheating of the hydraulics due to excessive strain
Burns, injuries caused by falling due to changes in functionality and damage to prosthetic com­
ponents
Do not touch overheated prosthetic components.
In case of changes in functionality, stop all activities and allow the overheated prosthetic
components to cool down.
In case of changes in functionality, reduce strain on the prosthetic component in order to
allow cooling.
In case of overheating or changes in functionality, have the prosthetic component inspected
for damage by a service centre authorised by the manufacturer.
Inform the patient.
NOTICE
Mechanical overload
Impaired functionality due to mechanical damage
Check the product for damage prior to each use.
Do not use the product if its functionality has been impaired.
Take any necessary measures (e.g. repair, replacement, inspection by the manufacturer's
customer service, etc.).
Inform the patient.
6.1 Using the vacuum pump
The vacuum pump is operated mechanically while walking with the prosthetic knee joint. A vacu­
um is created in the prosthetic socket during flexion while the medium being conveyed is
exhausted through the discharge valve during extension. An elevated vacuum is reached in the
sealed socket environment after several pump cycles. A vacuum can be generated even before
walking by flexing and extending the prosthetic knee joint while standing.
6.2 Flushing the vacuum pump
CAUTION
Improper cleaning
Skin irritation, formation of eczema or infections due to contamination with germs
Clean the product regularly.
Conclusion
flexion,
EBS unit too soft (pretension too
low)
EBS unit too firm (pretension too
high)
Measures
Turn the adjustment nut to the right
(+)
Turn the adjustment nut to the left
(–)
Ottobock | 23