Herunterladen Inhalt Inhalt Diese Seite drucken

Ex Classification Atex; Introduction; Rated Data - Mettler Toledo InPro 7250 Serie Bedienungsanleitung

Induktive leitfähigkeitsmesszellen
Vorschau ausblenden Andere Handbücher für InPro 7250 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10
2.7

Ex classification ATEX

2.7.1

Introduction

According to RL2014/34/EU
lnPro 725X/*/*/* conductivity sensors are devices
group ll, category 1/2G and according to RL 99/92/EG
(ATEX 137) may be used in zones 0/1 or 1/2 or 1 or
2 and gas groups llA, llB and llC that are potentially ex-
plosive due to combustible substances in above stated
temperature classes.
For use / installation, the requirements of EN 60079-14
must be observed.
According to RL2014/34/EU
lnPro 725X/*/*/* conductivity sensors are devices
group ll, category 1/2D and according to RL 99/92/EG
(ATEX 137) may also be used in zones 20/21 or 21 or
22 that contain combustible dusts.
The conductivity measuring circuit and temperature
measuring circuit are part of a common intrinsically safe
system and is for operation connected to a separately
certified transmitter.
The conductivity measuring circuit and temperature
measuring circuit as part of an intrinsically safe system is
isolated from the not-intrinsically safe electric circuits up
to a maximum rated voltage of 375 V and from grounded
parts up to a maximum rated voltage of 30 V.
The sensors can be installed with different process adapt-
ers (flange, threaded bushes) permanently in pipes or
tanks.
2.7.2

Rated data

Conductivity measuring circuit, and temperature meas-
uring circuit1:
In type of protection intrinsic safe Ex ia IIC, only for con-
nection to a certified intrinsically safe circuit.
Maximum values:
Ui ≤ 16 V, Ii ≤150 mA, Pi ≤155 mW
Li = 0 (the internal inductance is ineffective towards the
Ci = 9 00 pF (effective internal capacitance)
Note:
The above maximum values are each the total of all in-
dividual circuits of the associated intrinsically safe power
supply and transmitter.
1)
2)
InPro 7250 Series
Printed in Switzerland
outside)
For UK statutory Requirements SI 2016 No. 1107
For UK regulations 7 and 11 of the Dangerous Substances and
Explosive Atmospheres Regulations 2002 (DSEAR).
InPro 7250 Series
(ATEX 114) Appendix l,
1)
(ATEX 114) Appendix l,
1)
© 10 / 22 METTLER TOLEDO