Other Functions

11-1 Initialized Program (Factory Setting)
The setting value for each of the modes are stored in this product. These values can be initialized to the factory settings.
1) Turn the POWER switch ON while depressing the Foot Control button B.
2) "Clr" is displayed on the LCD Panel. While "Clr" is displayed, release the Foot Control button B and then depress the button D to start
initializing.
3) "Fin" is displayed on the LCD Panel when the initialization is complete. An alarm sound is heard and the product starts in the normal mode.
(Refer to the table below for the initial settings.)
Ultrasonic Mode selection
Power Level
Irrigation Mode
Irrigation Volume
11-2 Using the Tip Holder (for Ultrasonic Tip, option)
• U se the Tip Holder to store the Tips after removing them from the Handpiece.
• T he Tip Holder is autoclavable. Up to 5 Tips can be sterilized at the same time.
Rotate the Holder in the direction of the arrow so the Tips are set in the frame for
sterilization (Fig. 61).
11-3 Basic Water Level Setting
Basic Water Level can be adjusted depending on the supply water pressure in the
usage environment. If the Irrigation Volume is too large or too small even when
adjusted by the Irrigation Volume Keys, follow the procedure below to adjust the
Basic Water Level.
1) Turn the POWER switch of the Control Unit to ON.
2) Touch the Operation Mode Select Key to select
Powder Mode.
3) Touch and hold the Irrigation Mode Select Key for
2 seconds. Alarm sounds are heard and the LCD
panel switches to the Basic Water Level
adjustment mode.
4) Touch the Irrigation Volume Keys ( / ) to
change the Basic Water Level ("L1" - "L5"). Adjust
the Level as necessary.
5) Touch and hold the Irrigation Mode Select Key for
2 seconds to save the selected level to memory.
An alarm sound is heard and the selected level is
saved, and then the LCD Panel switches back to
normal mode.
NOTICE • Basic Water Level returns to "L4" when
this product is initialized to the factory
defaults. Set the Basic Water Level
again after initializing.
12

Periodical Maintenance Checks

Refer to the check sheet below and perform periodical maintenance checks every three months. Contact your Authorized NSK Dealer if any
abnormalities are found.
Points to Check
Ultrasonic Handpiece
Chamber Cap
Ultrasonic Mode
Perio
1
Bottle
10
Operate the Ultrasonic Handpiece to confirm there is no abnormal vibration, noise, heat, etc.
Also check that the LED lights and water flows.
Check that there are no large/deep scratches.
30
Powder Mode
Prophy Mode
1
(fixed to tap water)
10
Tip
Fig. 62 : Basic Water Level adjustment mode [Water Level 4]
Reference
Level
L5
L4 (Initial value)
L3
L2
L1
Details
Perio Mode
1
10
Tip Holder
Fig. 61
Water volume
Large
Small