Fault
Device does not operate
(LCD screen off)
Device does not operate
(LCD screen on)
Motor does not operate
Torque fault
Speed fault
No spray
Inadequate spray
Micro-motor does not
operate
Power fault
weak tip vibration
No ultrasonic function
Possible cause
Power cord not plugged in
properly.
Power switch in position O (OFF).
No power voltage.
Power receptacle fuse(s) blown.
Internal fuse blown
Transmission fault.
Motor or motor connections
problem.
Thermal cut-out.
Incorrect torque adjustment.
Wrong contra-angle.
Incorrect speed adjustment.
Wrong contra-angle.
Irrigation solution bag or bottle
empty.
Irrigation off.
Irrigation line clogged.
Line end piece on contra-angle
clogged.
Incorrect irrigation flow rate
adjustment.
Faulty micro-motor cord
connector electrical contacts.
Faulty micro-motor electrical
contacts.
Micro-motor cord wires cut.
Motor damaged.
Tip worn or distorted.
Incorrect power adjustment.
Incorrect use: incorrect angle of
attack or inadequate pressure on
the tooth.
Liquid or moisture present
between handpiece and cord.
Incorrect tip tightening.
Faulty connector contact.
Remedy
Check the electric socket.
Return to SATELEC after-sales.
Place the power switch in position I
(ON).
Call an electrician.
Replace the power receptacle fuses.
Return to SATELEC after-sales.
Switch the device OFF then ON again.
Return to SATELEC after-sales.
Check the motor cord connector.
Check that the cord connector is
properly plugged into the motor
connector of the device.
Return to SATELEC after-sales.
Allow the micro-motor to cool down.
Adjust the torque in accordance with
good dental practice.
Change the contra-angle.
Return to SATELEC after-sales.
Adjust the speed in accordance with
good dental practice.
Change the contra-angle.
Return to SATELEC after-sales.
Replace the bag or bottle of irrigation
solution.
Press the irrigation ON/OFF button.
Change irrigation line.
Clear the contra-angle line end piece.
Adjust the irrigation flow rate.
Clean the electrical contacts of the
micro-motor cord connectors.
Clean the electrical contacts of the
micro-motor.
Return to SATELEC after-sales to
replace the cord.
Return to SATELEC after-sales to
replace the motor.
Replace the tip.
Refer to the TipBook for the
necessary information.
Refer to the TipBook for the
necessary information.
Dry the electrical contacts
thoroughly.
Retighten the tip with the wrench.
Clean the connector contacts.