To reduce the risk of accidents, the precautions stipulated below must be taken:
Device users:
- Use of the I
C
2 is restricted solely to qualified, trained and competent dental health
MPLANT
ENTER
practitioners in the normal context of their work.
- If you have received this device by error, please contact the supplier so that it can be removed.
Interactions:
- The I
C
2 must not be used if the patient and/or the operator has a cardiac stimulator
MPLANT
ENTER
(pacemaker) or any other active implant (e.g. a cochlear implant).
- The device complies with applicable electromagnetic compatibility standards. The user should
nevertheless ensure that any potential electromagnetic interference does not cause an additional risk
(presence of radiofrequency emitters, electronic devices, etc.).
- The device is not designed to withstand shocks delivered by an electric defibrillator.
- Do not attempt to connect to the I
C
2 any other accessories than those supplied by
MPLANT
ENTER
SATELEC.
Electrical connection:
- Your I
C
2 must be connected to the electric power supply by a certified dental installation
MPLANT
ENTER
technician.
- Warning: To avoid any risk of electric shock, this device must be connected to an electrical power
supply with a protective ground.
- The electric supply to which the device is connected must comply with the standards in force in your
country.
- If loss of electrical power during use is likely to generate an unacceptable risk, the device must be
connected to a suitable power source (e.g. UPS).
Using the device:
- Do not use the device if it appears to be damaged or faulty.
- Turn the device off before unplugging the power cord.
- To unplug the power cord, grip the cord plug and hold the wall socket.
- Never use any other irrigation solution containers than those intended for suspension from the supplied
brackets.
- The device must only be used with bottles or bags of physiological saline or sterile water.
- The capacity of the irrigation solution containers used must not exceed one litre.
- When the device is not to be used for a long period of time, unplug the device from the electric
supply.
- Do not exert excessive force on the screen.
- Do not move I
C
2 during use.
MPLANT
ENTER
Environment:
- Do not cover the device or obstruct the ventilation vents.
- Do not immerse the device in liquid, and do not use it outdoors.
- Do not tilt the device at an angle greater than 5°.
- Do not place the device near a heat source.
- Make sure that the cords are not in a traffic path.
- The device must be stored in its original packaging in an appropriate and safe place.
- The device is not designed for use in the presence of anesthetic gases or any other flammable gas.
- Do not expose the device to water vapor, or splashes.
- Any condensation inside an electrical device is potentially dangerous.