Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroLab is intended for use in the electromagnetic environment specified below.
The customer or the user of the MicroLab should assure that it is used in such an
environment.
Immunity
Test
Conducted RF
3 Vrms
IEC61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC61000-4-3
80 MHz to
2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
IEC 60601
Compliance
test level
Level
Electromagnetic Environment -
Guidance
Portable and mobile RF
communications equipment should be
used no closer to any part of the
MicroLab, including any cables, than
the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance (d)
3 Vrms
d = 1.2√P
3 V/m
d = 1.2√P
d = 2.3√P
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Fields strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey,
should be less than the compliance
level in each frequency range.
Interference may occur in the vicinity
of equipment marked with the
following symbol:
90
80 MHz to 800 MHz
800 MHz to 2.5 GHz
a
b