Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The MicroLab is intended for use in the electromagnetic environment specified below.
The customer or the user of the MicroLab should assure that it is used in such an
environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations /
flicker emissions
IEC61000-3-3
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MicroLab is intended for use in the electromagnetic environment specified below.
The customer or the user of the MicroLab should assure that it is used in such an
environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient / burst
IEC61000-4-4
Surge
IEC61000-4-5
Compliance
Group 1
Group B
Class A
Complies
IEC 60601 test
level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Electromagnetic Environment -
Guidance
The MicroLab uses RF energy only for
its internal function. Therefore, its RF
emissions are very low and are not likely
to cause any interference in nearby
electronic equipment.
The Microlab is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes
Compliance
Level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
Input/output line
tests not
applicable (<3m)
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
88
Electromagnetic
Environment -
Guidance
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.