OtoRead Instruction for Use - English
Date: 2004-03-24
Do not immerse the unit in any fluids. See the operation manual for proper
cleaning procedures.
Do not drop or otherwise cause undue impact to this device. If the instrument
is dropped or otherwise damaged, return it to the manufacturer for repair
and/or calibration. Do not use the instrument if any damage is suspected.
Use and store the instrument indoors only. Do not use this instrument or its
accessories in temperatures below 4°C /40°F or above 38°C /100°F, or in
relative humidity of more than 90%.
Do not attempt to open or service the instrument. Return the instrument to the
manufacturer for all service. Opening the instrument case will void the
warranty.
Do not operate the printer if the power supply has a damaged cord or plug.
See the instructions below.
Be sure to use only stimulation intensities, which will be acceptable for the
patient.
The transducers (headphones, bone conductor, etc.) supplied with the
instrument are calibrated to this instrument - exchange of transducers require
a recalibration.
Please note that the CE Marking is only legal if this Instruction is translated
into the national language of the user no later than at the delivery to him, if the
national legislation demands a text in the national language according to MDD
article 4.4.
Although the instrument fulfils the relevant EMC requirements precautions
should be taken to avoid unnecessary exposure to electromagnetic fields, e.g.
from mobile phones etc. If the device is used adjacent to other equipment it
must be observed that no mutual disturbance appears.
Disposal of batteries must be made according to national regulations.
If this apparatus is connected to one or more other devices with medical CE
marking, to make up a system or pack, the CE marking is only valid also for
the combination if the supplier has issued a declaration stating that the
requirements in the Medical Device Directive article 12 are fulfilled for the
combination.
Printer AC Power Source Precautions
The printer AC power source converts main voltage (120V AC or 220V AC) to
a low voltage (7V DC) for powering the printer. Use only the power source
supplied for this purpose.
Notice - This product is for indoor use only. Do not expose to water or
excessive dust.
Notice - This product is not suitable for operation in the presence of
flammable anesthetic mixture with air or with nitrous oxide.