Export the waveform into a file in specified format according to current record length. You can choose
bin, txt, csv, xls as the format. You can also click
icon on the user interface to save it in the
format specified in menu.
Self Cal
The self-calibration can make the oscilloscope reach the optimum condition rapidly to obtain the most
accurate measurement value. You can carry out this application program at any time. This program
must be executed whenever the change of ambient temperature is 5°C or over.
Before performing a self-calibration, disconnect all probes or wires from the input connector. After
everything is ready, click "Self Cal".
Default
Click "Default" to call out the factory settings. You can also click icon on the user interface.
Help
Open this help document. You can press F1 as shortcut key.
20. Network
The oscilloscope could be connected to a PC via LAN port. For more details, see
22. Use LAN
Port.
Tips Window
Choose a label among "Channel", "Capture&Period" or "Trigger" to view the guide for the new users.
Click anywhere in the help window to turn page.
If you do not want to display the guide automatically when running the software, check "Don't show
again".
About
Display Version, Serial Number and Company website.
MULTI I/O
Control the function of the port 5 in Ports of the Oscilloscope.
Trigger In: Input the trigger signal synchronously
Trigger Out: Output the trigger signal synchronously
Pass/Fail: Output high level when pass, output low level when fail.