Herunterladen Inhalt Inhalt Diese Seite drucken

Specifications - PeakTech 1075 Bedienungsanleitung

Digital-multimeter
Vorschau ausblenden Andere Handbücher für 1075:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4. Specifications

Accuracies are: (% of reading + no. of digits) guaranteed for 1 year,
23° C ± 5° C, less than 75 % R.H.
Note:
Accuracy specifications consist of two elements:
(% reading) – This is the accuracy of the measurement circuit.
(+ digits) – This is the accuracy of the A/D-converter.
DC Voltage
Range
Accuracy
200 mV
2000 mV
± 0.5 % + 2 digit
20
V
200
V
± 0.8 % + 2 digit
600
V
Input Impedance:
Overload protection:
AC Voltage
Range
Accuracy
200 V
± 1.2 % + 10 digit
600 V
Input Impedance:
Frequency range:
Overload protection:
Indication:
Resolution
0,1 mV
1
mV
10
mV
100
mV
1
V
> 1 M on all ranges
600 V DC or peak AC
on all ranges,
rms
in 200 mV range 200 V
rms
Resolution
100 mV
1 V
> 1M on all ranges
50 Hz to 450 Hz (>60 Hz unspecified)
600 V DC or AC
in all ranges
rms
Average (rms of sine wave)
-23-

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis