6.4 Hysteresis
If the level is fluctuating around the set value (e.g., wave movement during filling), the hysteresis
keeps the switching state of the outputs stable. When the level is increasing, the output switches
when the respective switching point (SP) is reached; if the level sinks again, the output switches back
only after the reset point (RP) has been reached.
6.5 Application notes
The error signal of the sensor is analogous to the cable breakage control.
Dry run protection / empty signal
38
8014638/ZQ14/2017-08-25 © SICK AG • Level sensors • Germany • Subject to change without notice
6 Operation
Description
This function makes it possible to save diagnosic data in the device.
This parameter describes the quality of the measurement signal by using two
values; signal strength and signal interference.
Signal strength:
> 100 very good signal quality
< 20 poor signal quality
Signal interference:
> 80 strong signal interference
Entering the password to gain access into protected menus:
Password 000537 (LFP on cell phone keyboard / L=5 / F=3 /P=7)
With a incorrect password or power up you can logg the expert menue.