Herunterladen Inhalt Inhalt Diese Seite drucken

Eyepieces - Leica DMI3000B Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für DMI3000B:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Depending on the configuration, the screen will
now display
- the active port (right or left) and
- the percentage of light going to the port
(100%, 80%, 50%).
Optional Leica DMI 6000 B:
The bottom port selection function can be as-
signed to one of the variable function buttons on
the stand or the SmartMove.
The top port can only be selected manually.
Select ports via
fixed function buttons on stand (side
ports)
variable function buttons on stand and
SmartMove (bottom port)
manual action (top port)

8.8 Eyepieces

Note:
The eyepiece's aperture protector must be re-
moved or folded back, during microscopy while
wearing eyeglasses. We recommend removing bi-
focals and spectacles with progressive-addition
lenses when using the microscope.
• For the adjustable tubes with documentation
output, choose the 100% VIS position.
Eyepieces with inlaid reticle
• Focus the reticle by adjusting the eyelens.
• Focus on the object through this eyepiece.
• Then, close that eye and focus on the
object by adjusting only the second ocular.
Correction for Vision Problems
• With your right eye, look through the right
eyepiece and bring the specimen into sharp
focus.
• Then, with your left eye, view the same speci-
men and rotate the left eyepiece tube until the
object is brought into sharp focus. Do not
change the Z position in the process!
8. Operation
97

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Dmi4000bDmi6000b

Inhaltsverzeichnis