Herunterladen Inhalt Inhalt Diese Seite drucken
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

16
Tenupol-3
Instruction Manual

2.7. Operation Parameters

2.7.1.
Adjust Mode
For basic adjustment of flow rate, the apparatus should be in "Adjust Mode".
In this mode the normal switch-off security systems and the polishing circuit
are not in operation.
Set the apparatus in "Adjust Mode" as follows:
Replace the rear jet holder by the jet holder with the ascending tube.
Remove the thermometer from the polishing cell.
Place the return tube in the thermometer pocket and make sure that the
ascending tube goes inside the return tube.
Place a specimen holder, with a specimen inserted, in the polishing
cell.
Adjust the flow rate to minimum (turn the flow rate knob counter--
clockwise).
Switch off the power switch.
With the Start/Stop-button activated, turn on the power switch.
Release the Start/Stop-button.
Now the motor is working with the low flow rate and the light of the
polishing diode is yellow. Dependent on other factors the light of the
Start/Stop-diode may be either red or green.
Disconnect "Adjust Mode" in the following way:
Disconnect "Adjust Mode" either by pushing the Start/Stop-button or by
switching off the main switch.
Remove the specimen holder from the polishing cell.
Replace the jet holder with ascending tube by the holder with cathode.
Remove the return tube and insert the thermometer.
Clean the jet holder with the ascending tube
and the return tube with alcohol after use.
NB

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis