08/2016
1009864-02A
Water and cooling conditions
Permissible water pressure (water inlet): 8 bar
Maximum coolant temperature: 16°C
The device is not to be used outdoors.
The device is not suitable for operation in mobile facilities (ships, buses, trains
or oil platform). The unit has to be adapted in order to guarantee trouble-free
operation when used in mobile facilities.
Adaptation of the unit is required for operating altitudes greater than 2000 m
above sea level in order to guarantee trouble-free operation.
Please contact the customer service department without fail before putting the
device into operation - see page 2.
1.5.5
Commissioning
Initial commissioning will be undertaken by the manufacturer's own service
personnel or by an authorized service company. The device will be handed
over ready for use.
If the device has not been used for a longer period of time then it must first be
inspected by our customer service staff or by another authorized company
before being put back into operation again.
The device is only to be used if it is working correctly!
None of the safety devices (e.g. switch) or safety warnings are to be removed
or deactivated as this would impair the safe operation of the device!
The device must never be used if it is empty.
A water cooling system must be connected to the device if it is to be used for
more than 1 hour per day at maximum pressure. The high pressure hoses
must not be removed. Please contact the customer service department with-
out fail before putting the device into operation – see page 2.
The device may only be operated with the high-pressure hoses included in
the scope of delivery.
A water shut-off valve must be used.
High frequency mobile communication systems can interfere with the way the
device functions.
Heat therapy devices (microwave therapy devices, etc.) installed in the vicinity
may lead to electromagnetic interference. Contact our customer service de-
partment if this happens – see page 2.
Venting
The tank must be vented 3 days after commissioning or after it has been
initially filled – see page 39. After this, venting is unnecessary unless leak-
ages occur and the tank has to be topped up.
The device may no longer be used in the event of water loss (puddles on the
floor). Please inform our customer service department – see page 2.
Safety instructions and warnings
Operating instructions – 13/46