Herunterladen Inhalt Inhalt Diese Seite drucken

Rotor Detection - Hettich EBA 280 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für EBA 280:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

18

Rotor detection

The centrifuge is equipped with automatic rotor detection.
Rotor detection is carried out after starting each centrifugation run.
If the rotor code of a rotor is read in for the first time by the rotor detection, the centrifugation run then stops
automatically after rotor detection. The rotor code (R) and the maximum speed (maxRPM) of the rotor permitted for
this centrifuge are displayed.
Example:
PROG
RPM
t/min:s
After the rotor comes to a standstill (the left LED in the
maximum permissible number of run cycles is displayed.
Example:
PROG
RPM
t/min:s
Now, the maximum permissible number of run cycles of the rotor/hangers must be set (see the "Cycle counter"
chapter).
When using a rotor whose rotor code has already been read in by the rotor detection, after starting the centrifugation
run, the maximum speed (ROTOR MAX) of the rotor permissible for this centrifuge is briefly displayed.
Example:
PROG
RPM
t/min:s
If the set speed is greater than the maximum speed of the rotor permissible for this centrifuge, the
centrifugation run is aborted after rotor detection and the following displays alternately appear:
Example:
PROG
RPM
In this case, the speed on the display must be checked and adapted to the maximum rotor speed.
t/min:s
PROG
RPM
STOP
key is illuminated), press any key. Afterwards, the
OPEN
t/min:s
EN