Message / fault
CTRL ERROR
CRC ERROR
COM ERROR
FC ERROR
61.1 – 61.21,
61.64 – 61.142
TACHO ERR
FC ERROR
FC ERROR
VERS. ERR
82/165
Cause
22 – 25.4
Error / defect electronics
27, 27.1
31 – 36
60,
61.22
Speed measurement error
61.23
61.153
Error / defect electronics
61.154
Invalid machine version
Remedy
Perform a MAINS RESET when the
rotor has been stationary.
The device must not be switched off
as long as the "wait" message is
being displayed. After the message
"wait ...!" is no longer displayed,
carry out a MAINS RESET.
Perform a MAINS RESET.
Check the loading of the rotor, see
chapter "Loading the rotor".
Repeat the centrifugation run.
Perform a MAINS RESET.